Please oder LMG in combination of a sensor and a display unit.

 

■Sensor specification

 MODEL LMG 027 LMG 127 LMG 307 LMG 607 LMG 807 LMG 1507
Dimensional drawing PDF PDF PDF PDF PDF PDF
Light source Visible Diode Laser (Wave Length : 655nm)
Measuring range(mm) 0.01 – 2 0.05 – 12 0.2 – 30 0.5 – 60 0.5 – 80 0.7 – 150
Resolution (micron) 0.01 0.1 1
(inch) 0.000001 0.00001 0.0001
Measuring Vertical(mm) 2 12 30 60 80 150
area Horizontal(mm) 0 *2 4 (2) *3 10 30 40 20
Accuracy*1  Reproducibility(µm) -0.5 -1 -2 -4 -5 ±12(±5)*4
  Linearity (µm) -0.3 -0.5 -1 -2.5 -4 ±8(±4)*4
  Repeatability(µm) -0.03 -0.04 -0.1 -0.35 -0.6 ±1.3(±1)*4
Scanning freq. 3600 / s
Scanning speed (m/s) 85 226 355 707 990 1130
Operation temp (deg.C) 0 – 45
Weight(kg)  Trasmitter 2 2.5 2.7
Receiver
Base plate  - 
Total Weight(kg) 2 2.5 2.7

*1 At ambient temparature of 20+-5 deg.C. About diameter measuring value of cylindrical gauges.

Reproducibility
Deviations of value measured in the length X and breadth Y area around the measurement central position from the value measured at the central position (X is 70% of the vertical length of its measuring area and Y is 50% of the horizontal length of it).
Linearity
Errors of different size objects measured at the measuring area center.

Repeatability
The value of +-2σ while measuring a gauge repeatedly in 0.25 second (896 average times) intervals at the measuring area center (σ: a standard deviation)

*2 At horizontal center only
*3 ( ) : For from 0.05mm to 0.1mm measuring objects
*4 ( ) : For up to 80mm diameters
*5 ( ) : At lower measuring area

 

■Display unit

 

Specifications D7 D7-M
Number of connecting sensors 1 2
Dimensional drawing PDF
Display Two measuring values can be displayed at the same time.
Display unit Selectable between mm and inch
Display interval 0.28ms to 4.4s Selectable : 5 kinds of interval registered
Nominal value Pre-settable for each measuring mode
Tolerance value Pre-settable each Hi tolerance and Lo tolerance to each displayed values
Measuring mode Selectable between 2 point diameters and 3 point gaps.
Data processing Mean value, Max. value, Min.value, Max-Min. value, Moving average, etc.
Data output RS-232C, Analog±10V(Option :GP-IB, Centronics, BCD)
Alarm Hi tolerance over, Lo tolerance over, whithin tolerance: Lamp on relay output
Others Measurement by input signal. Offset. Zeroset, Calibration etc.
Weight 4kg
Power Source AC85V – 264V 50/60Hz 30VA

 

Standard accessories

・Connecting cable between transmitter and receiver ·········· 1
・Connecting cable between sensor and display(4m) ·········· 1
・Power cable (2m) ························································· 1
・Operation Manual ························································· 1

 

Options

・Transparent object measuring function
・Multi-segments measuring function of outer diameter and gap.
・BCD,Centronics & GP-IB interface
・Dustproof hood(for protection against dust)
・Standard gauges
・Design & production of measureing line